Optimization of ordered subset expectation maximization parameters for image reconstruction in Tc-99m methoxyisobutylisonitrile myocardial perfusion SPECT and comparison with corresponding filtered back projection-reconstructed images


Pankaj Dheer, Priyanka Gupta, Sameer Kamalakar Taywade, Averilicia Passah, Anil Kumar Pandey, Chetan Patel 
 Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi, India

Correspondence Address:
Dr. Chetan Patel
Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi - 110 029
India

Purpose of the Study: To establish the most appropriate ordered subset expectation maximization (OSEM) parameters for image reconstruction in Tc-99m methoxyisobutylisonitrile (MIBI) myocardial perfusion SPECT (MPS) and comparison with corresponding filtered back projection (FBP)-reconstructed images. Methods: A total of 99 stress–rest MPS studies (47 normal and 52 abnormal) were retrospectively analyzed using 16 different combinations of iterations and subsets. Images were reconstructed both with and without postreconstruction Butterworth filter (cutoff frequency and order for stress: 0.4 and 10 and for rest: 0.52 and 5, respectively) for each combination. A total of 3168 images were evaluated qualitatively by two nuclear medicine physicians on a scoring scale of 1–4. Best visual quality image iteration–subset combination was determined for each patient both with and without Butterworth filter and was further compared with FBP-reconstructed image. The interobserver agreement was obtained using kappa statistics. Results: The best quality images were obtained using a combination of four iterations and six subsets for both with and without Butterworth filter. The value of kappa for interobserver agreement for OSEM images with Butterworth filter was 0.570 and for OSEM images without Butterworth filter was 0.857. On comparison, FBP images were better than OSEM-reconstructed images without Butterworth filter (P < 0.0001 calculated using Fisher's exact test) with substantial agreement (kappa = 0.628). However, OSEM-reconstructed images with Butterworth filter were better than FBP images and showed moderate agreement (kappa = 0.486). Conclusion: The most appropriate OSEM reconstruction parameter in Tc-99m MIBI MPS is 4-iteration and 6-subset combination. FBP-reconstructed images were better than the images reconstructed with OSEM without postreconstruction Butterworth filter. However, OSEM-reconstructed image with Butterworth filter was better than FBP images.


How to cite this article:
Dheer P, Gupta P, Taywade SK, Passah A, Pandey AK, Patel C. Optimization of ordered subset expectation maximization parameters for image reconstruction in Tc-99m methoxyisobutylisonitrile myocardial perfusion SPECT and comparison with corresponding filtered back projection-reconstructed images.Indian J Nucl Med 2021;36:14-20


How to cite this URL:
Dheer P, Gupta P, Taywade SK, Passah A, Pandey AK, Patel C. Optimization of ordered subset expectation maximization parameters for image reconstruction in Tc-99m methoxyisobutylisonitrile myocardial perfusion SPECT and comparison with corresponding filtered back projection-reconstructed images. Indian J Nucl Med [serial online] 2021 [cited 2021 Apr 22 ];36:14-20
Available from: https://www.ijnm.in/article.asp?issn=0972-3919;year=2021;volume=36;issue=1;spage=14;epage=20;aulast=Dheer;type=0