CASE REPORT
Year : 2021  |  Volume : 36  |  Issue : 2  |  Page : 179-182  

Importance of understanding and analyzing daily quality assurance test of positron emission tomography/computed tomography equipment in minimizing the downtime of equipment in remote places


1 Department of Nuclear Medicine, Kailash Cancer Hospital and Research Centre, Vadodara, Gujarat, India
2 Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital; Department of Nuclear Medicine, Homi Bhabha National Institute, Mumbai, Maharashtra, India

Date of Submission21-Aug-2020
Date of Decision25-Sep-2020
Date of Acceptance29-Sep-2020
Date of Web Publication21-Jun-2021

Correspondence Address:
Mr. Sachin Tayal
Department of Nuclear Medicine, Kailash Cancer Hospital and Research Centre, Muni Seva Ashram, Goraj, Vadodara, Gujarat
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijnm.IJNM_196_20

Rights and Permissions
   Abstract 


This article briefly describes the event of a defective detector block in a daily quality assurance scan/blank scan and insists on implementing guidelines to scan or not to scan in such a scenario. The nuclear medicine physicist should have a clear understanding of the blank scan graph, which shall help rectify the right cause of problem and give confidence to the physician in reporting the acquired study. A routine blank scan in positron emission tomography signifies various parameters of the crystal (coincidence count rate, single count rate, dead time, and coincidence time along with energy response) and in some respect is analogous to the daily uniformity flood image for gamma cameras, providing an overall assessment of detector response. We encountered a bad detector block in our routine quality assurance scan/blank scan and analyzed the root cause behind such an error which was finally restored to normalcy by replacing the defected part with a new one and an error-free blank scan was established. The analysis was carried out by performing various possible checks and discussing the issue with service engineer to help identify the defects much before service engineer actually arrived in our department. This allowed us to take the correct decision and enabled us to get the scanner repaired faster. Hence, a good understanding of the daily quality control test and proper analysis of the same may result in swift decision-making and faster repair of equipment leading to minimal disruption in the clinical workflow as well as avoidance of suboptimal scanning leading to the wrong diagnosis.

Keywords: Block detectors, crystal map, photomultiplier tube tuning, positron emission tomography


How to cite this article:
Tayal S, Ali A, Kumar V, Jha AK, Gandhi A. Importance of understanding and analyzing daily quality assurance test of positron emission tomography/computed tomography equipment in minimizing the downtime of equipment in remote places. Indian J Nucl Med 2021;36:179-82

How to cite this URL:
Tayal S, Ali A, Kumar V, Jha AK, Gandhi A. Importance of understanding and analyzing daily quality assurance test of positron emission tomography/computed tomography equipment in minimizing the downtime of equipment in remote places. Indian J Nucl Med [serial online] 2021 [cited 2021 Jul 30];36:179-82. Available from: https://www.ijnm.in/text.asp?2021/36/2/179/318872




   Introduction Top


With the rapid growth of cancer worldwide, positron emission tomography (PET) has become a major tool for helping in cancer staging and response evaluation. This has put a greater responsibility on the Nuclear medicine physicist for obtaining accurate results from the system as he/she should be confident enough with the scanner's performance. Hence, a quality assurance program for the scanner should be established to ensure that the performance of a procedure or instrument is within a predefined acceptable range.[1],[2],[3],[4] Second, a guideline should be established to follow in case of various errors encountered as it shall help in the smooth functioning of the department. It is the rationale that in a busy department, performing all possible quality assurance test might be difficult but there should be a minimum requirement for the daily quality assurance (DQA) scan/blank scan, to test the scanner under normal conditions and periodic recalibrations be carried out to guarantee correct normalization.[2],[3] The DQA has unparalleled advantages as it is less time-consuming and can help in the evaluation of hardware-related artifact too.[5] It helps display in graphical representation various parameters of the crystal in PET (coincidence count rate, single count rate, dead time, coincidence time, and energy response). A PET system is made of an array of scintillation crystals, consisting of multiple blocks, and optically coupled to four or more single-anode photomultiplier tube (PMT) wherein a block damage of any scintillator can occur during the routine process without any warning. Hence, the knowledge of the malfunctioning block and its cause can be useful to prevent the cancellation of patients. Second, due to the quadrant sharing of PMT, the gain stability of all the PMT's is also a factor to be taken care of for the stable functioning of the system.[6] These defective detector blocks reading can cause serious image artifacts.[7] Some minor artifacts do not have a final impact on the diagnosis and can easily be recognized by an expert, but few artifacts may bring serious impacts on the image quality and finally lead to the wrong diagnosis.[8] One such case of defective crystal was observed in our department.


   Case Report Top


As per the routine practice, we were acquiring a DQA scan, with the inbuilt Ge-68 rod source of our PET-computed tomography (CT) scanner (Discovery STE, GE Medical System, USA) before injecting the patients with F18-Fluoro Deoxy Glucose for PET-CT study. On completion of the DQA scan, we observed a black spot at module 7 of block 6 [[Figure 1]a arrow mark]. From our previous experience; we could understand that it could be due to reasons such as malfunctioning of detector, temperature-humidity variation, or Cassette Electronic Module board (CEM board) error. We started analyzing each point to rule out the cause of a bad block detector. The temperature-humidity factor was ruled out as the gantry and console room temperature and humidity were well within the prescribed limit (20°C ± 2°C; 45 ± 5%) and stable for quite a few days, neither was there any error in the CEM board because an error in CEM board causes malfunctioning of a whole module and not a single block of a module. Hence, we shifted our analysis to detector malfunctioning, which could have been due to a nonfunctional detector or having a substantial lower sensitivity than that of the other detectors.[7],[8] We looked up the two-dimensional (2D) generated crystal decoding map of module 7, block 6. It clearly showed the distorted map with uneven/loss of distribution of signal [Figure 2]a. Meanwhile, we contacted a service engineer and briefed the situation and discussed with him. This prompted us to proceed further for analyzing the PMT gain/peak of the supporting module's block. The peak was completely derailed; which was leading to poor outcomes of the detector block [Figure 2]b. We discussed the outcome of PMT gain with service engineer and concluded the problem in one of the PMT in the 7th module. Service engineer ordered the detector module from a nearby warehouse. We also discussed this issue with our Nuclear medicine physician to decide whether to inject patients or not under such a situation. After the discussion, we performed a dummy scan to see whether the system was generating raw data for both attenuation correction and nonattenuation correction files and whether the fusion was taking place with the selected CT data. On its assurance, we along with the Nuclear medicine physician reanalyzed the DQA graph and since the region of a defect was confined to < 1% of total crystals, we decided to proceed ahead with selected routine PET-CT studies where accuracy in standardized uptake value (SUV) quantification was not really mandatory as per Nuclear medicine physician and rescheduled the cases in which SUV quantification was of prime importance: Like early response evaluation after therapy and at the same time registered a service call as per departmental protocol to GE service support. The images acquired were clearly interpretable with minimal degradation in the qualitative analysis [[Figure 3]a,[Figure 3]b,[Figure 3]c arrow mark]. By the time, we completed a study of selected few patients the service engineer arrived. After scrutinizing thoroughly through the software and error log along with inspection of completed studies and internal components/hardware, we again proceeded first with system calibration, which makes coarse and fine gain adjustments to the PMT, but we had the same graph of DQA.[11] Late in the evening part ordered also arrived and replaced it successfully in the early morning the next day. The detector module is shown in [Figure 2]c and [Figure 2]d. On installation and updating the PMT gain, we performed the DQA scan. On review by the service engineer and physicist in the department, DQA was found to be normal [Figure 1]b. On complete satisfaction, we performed a clinical PET-CT study successfully and there were no artifacts in the PET images [Figure 3]d,[Figure 3]e,[Figure 3]f.
Figure 1: Daily Quality assurance graph, (a) Shows defective graph arrow sign (b) shows a normal graph

Click here to view
Figure 2: Detailed graph at module level; (a) graph shows loss of signal on two-dimensional generated crystal decoding map of module 7, block 6, (b) graph shows loss of signal in energy peaking map of module 7, block 6, (c) figure shows the upper part of the detector module, and (d) figure shows the lower part of a detector module

Click here to view
Figure 3: Clinical scan with a defective module and after replacement; (a-c) arrows show photo deficient area on a clinical scan performed with defective detectors. (d-f) No photo de-ficient area in the clinical image performed after replacement of the detector module

Click here to view



   Discussion Top


Our system is a 35 (0–34) module Bismuth Germanate BGO crystal PET system, with each module having 8 blocks (0to7), wherein each block is having 6 rows (0–5) and 8 columns (0–7) of crystals, making a total array of 13,440 crystals (35 × 8 × 48). This array of a crystal is optically coupled to eight single-anodes PMT as shown in [Figure 2]c. The decoding of crystal position relies highly on the relative gain stability of all the PMT. Any drift in gain of single PMT relative to other can cause distortion in 2D crystal encoding map and assign wrong crystal of a gamma hit.[12] Various factors can lead to PMT tube gain change, such as room temperature, patient load, short term or long-term radiation exposure, and time, leading to degraded image resolution and quality in a PET camera. It has been observed that long-term PMT gain drop is not due to temperature, but due to the usage (light load) of the PMT, second, a short-term drift could be caused by idling of camera for 1–3 days, which can temporarily desensitize its PMT by 3%–12%, but a higher radiation load for the camera, especially after a few days of Idling can de-sensitze its PMT by 4%–17%.[12] A long-term variation could be recovered by system-wide tuning every few weeks or months in current commercial PET cameras, but it cannot overcome short-term variation. In our case, we faced the error in the middle of the week, hence the short-term drift was ruled out and the only factor which could lead to PMT drift was its long usage and routine wear and tear. Earlier too such a scenario has been experienced by the various users of modality, wherein. One of the studies performed by ElHami et al. has shown a mean change in phantom SUV max was-2%(range,-6% to + 3%) in the presence of a single defective block detector and-3% (range,-11% to + 7%) in the presence of two defective block detectors respectively. Moreover, the routine patients scan could be performed keeping the reading physician aware of the detector failure.[13] Second study performed by Samiee et al. have concluded that for the case of one to two defective detectors in a system, reasonably quantitative clinical PET imaging can be performed and it might be possible to acquire useable images.[14] Zito et al., in their study, they demonstrated that nonuniformity parameters did not significantly change whenever assessed on original or an altered data having one block or two axial blocks (faulty analog board) with zero sensitivity.[15] In other study Jha et al. have shown the reduction of artifacts using the sinogram correction technique.[16] We were partly lucky to have the error confine to single-crystal block, avoiding the complete breakdown of the system, second a good understanding of the internal components and established guideline/flow chart to investigate the outcome of each part through its graphical representation can help us conclude to diagnose the right cause/origin of error much earlier than the arrival of service engineer. Such practice guidelines of understanding the cause and taking the decision in consensus with the reporting physician being aware of the current scenario and its limitation can help plan patients accordingly for optimal utilization of the modality in such scenarios and reduce the turnaround time to breakdown.


   Conclusion Top


Good understanding of the DQA test by qualified nuclear medicine physicist and proper analysis of the same may result in swift decision-making and faster repair of equipment leading to minimal disruption in the clinical work as well as avoidance of suboptimal scanning leading to the wrong diagnosis.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: A brief review. J Nucl Med 2008;49:1114-31.  Back to cited text no. 1
    
2.
Keim P. An overview of PET quality assurance procedures: Part 1. J Nucl Med Technol 1994;22:27-34.  Back to cited text no. 2
    
3.
Jha AK, Mithun S, Singh AM, Purandare NC, Shah S, Agrawal A, et al. 18-Month performance assessment of gemini TF 16 PET/CT system in a high-volume department. J Nucl Med Technol 2016;44:36-41.  Back to cited text no. 3
    
4.
Jha AK, Mithun S, Puranik AD, Purandare NC, Shah S, Agrawal A, et al. Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012. World J Nucl Med 2019;18:351-60.  Back to cited text no. 4
[PUBMED]  [Full text]  
5.
Tayal S, Gandhi AC, Ali A. Significance of daily quality assurance scan in hardware artifact evaluation. Indian J Nucl Med 2018;33:351-4.  Back to cited text no. 5
[PUBMED]  [Full text]  
6.
Li H, Liu Y, Xing T, Wang Y, Uribe J, Baghaei H, et al. An instantaneous photomultiplier tube gain-tuning method for PET or gamma camera detectors using an LED network. IEEE Trans Nucl Sci 2005;52:1295-9.  Back to cited text no. 6
    
7.
Buchert R, Bohuslavizki KH, Mester J, Clausen M. Quality assurance in PET: Evaluation of the clinical relevance of detector defects. J Nucl Med 1999;40:1657-65.  Back to cited text no. 7
    
8.
Hsieh J. Image artifacts: Appearances, causes and corrections. In: Computed Tomography: Principles, Design, Artifacts and Recent Advances. Bellingham, Wash: SPIE Press; 2003. p. 167-240.  Back to cited text no. 8
    
9.
Cherry SR, Sorenson J, Phelps ME, Methé BM. Physics in Nuclear Medicine. Vol. 31, Med Phys 2004;31:2370-71.  Back to cited text no. 9
    
10.
Zanzonico P. Positron emission tomography: A review of basic principles, scanner design and performance, and current systems. Semin Nucl Med 2004;34:87-111.  Back to cited text no. 10
    
11.
Keim P. An overview of PET quality assurance procedures: Part 2. J Nucl Med Technol 1994;22:27-34.  Back to cited text no. 11
    
12.
Uribe J, Li H, Baghaei H, Aykac M, Wang Y, Liu Y, et al. Effect of photomultiplier gain-drift and radiation exposure on 2D-map decoding of detector arrays used in positron emission tomography. IEEE Nucl Sci Symp Med Imaging Conf 2002;4:1960-4.  Back to cited text no. 12
    
13.
Elhami E, Samiee M, Demeter S, Leslie WD, Goertzen AL. On the significance of defective block detectors in clinical (18) F-FDG PET/CT imaging. Mol Imaging Biol 2011;13:265-74.  Back to cited text no. 13
    
14.
Samiee, Maryam and A. L. Goertzen. “Quantifying the effects of defective block detectors in a 3D whole body pet camera.” 2007 IEEE Nuclear Science Symposium Conference Record 6 (2007): 4258-4261.  Back to cited text no. 14
    
15.
Zito F, De Bernardi E, Schiavini M, Canzi C, Voltini F, Agosteo S, et al. Analysis of different detector and electronics defects on F18-FDG images. Nucl Instrum Meth A 2007;571:493-7.  Back to cited text no. 15
    
16.
Jha AK, Purandare NC, Shah S, Agrawal A, Puranik AD, Rangarajan V. PET reconstruction artifact can be minimized by using sinogram correction and filtered back-projection technique. Indian J Radiol Imaging 2014;24:103-6.  Back to cited text no. 16
[PUBMED]  [Full text]  


    Figures

  [Figure 1], [Figure 2], [Figure 3]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Case Report
   Discussion
   Conclusion
    References
    Article Figures

 Article Access Statistics
    Viewed340    
    Printed0    
    Emailed0    
    PDF Downloaded36    
    Comments [Add]    

Recommend this journal