ORIGINAL ARTICLE
Year : 2020  |  Volume : 35  |  Issue : 4  |  Page : 310-314

Assessment of radiation dose delivered and volume measurement by low- and high-dose diagnostic computed tomography: Anthropomorphic liver phantom study


Department of Nuclear Medicine, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India

Correspondence Address:
Dr. Sanjay Gambhir
Department of Nuclear Medicine, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, Uttar Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijnm.IJNM_44_20

Rights and Permissions

Aim: Liver volume measurement is a mandatory test before measure liver surgeries and transplantation. We aimed a study on the difference in volume measurement and radiation dose to an anthropomorphic liver phantom using high-dose and low-dose diagnostic computed tomography (CT). Materials and Methods: Several measurements of the manual total volume measurement done on an anthropomorphic liver phantom mounted with thermoluminescent dosimeter. We exposed the phantom with diagnostic CT, low-dose CT, and a low-dose CT with copper filter. Results: Phantom underwent ten scanning for each exposure. There was no significant difference in the total volume measurement in comparison to the phantom volume. The volume of phantom measured by low-dose CT, low-dose CT with copper phantom, and high-dose CT were 1869 ± 18 cm3, 1852 ± 24 cm3, and 1908 ± 12 cm3, (P = 0.3), respectively. However, the radiation dose delivered was significantly different (1.54 mGy, 0.77 mGy, and 5.84 mGy [P = 0.001], respectively). Conclusion: Total liver volume measurement provides essential clinical information in several clinical conditions. We recommended that the volume measured by a low-dose CT has an excellent correlation with the diagnostic quality CT and should be a routine in the routine clinical practice. CT volumetry achieves the same result while using very less radiation exposure. It may also be used with functional imaging to give complete information.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed112    
    Printed2    
    Emailed0    
    PDF Downloaded20    
    Comments [Add]    

Recommend this journal