ORIGINAL ARTICLE
Year : 2019  |  Volume : 34  |  Issue : 3  |  Page : 201-204  

Comparison of image quality of different radionuclides technetium-99m, samarium-153, and iodine-123


1 Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco
2 Department of Medical Physics, Academic Hospital of Udine, Udine, Italy

Date of Web Publication20-Jun-2019

Correspondence Address:
Dr. Youssef Bouzekraoui
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat
Morocco
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijnm.IJNM_42_19

Rights and Permissions
   Abstract 


Introduction: The choice of the radionuclide has a key role in nuclear medicine which appearing the lowest scatter fraction. In addition, the presence of penetrated and scattered photons from collimator in single-photon emission computed tomography images degrades resolution and contrast. Thus, image quality depends on sensitivity and resolution of the collimator–detector system. The goal of this study was to compare the image quality that can be achieved by three radionuclides: technetium-99 m (Tc-99 m), iodine-123 (I-123), and samarium-153 (Sm-153). Materials and Methods: Tc-99 m and Sm-153 were imaged with low-energy high resolution (LEHR) collimator, while I-123 was imaged with medium-energy (ME) collimator. We modeled the Siemens Symbia Medical system using Monte Carlo simulation code SIMIND. The imaging characteristics of each radionuclide were investigated by simulated data: point spread function, sensitivity (Cps/MBq) and geometric, penetration and scattering distribution. Results: Tc-99 m and Sm-153 give best and results with LEHR collimator for spatial resolution (full width at half maximum [FWHM] = 3.19 mm; full width at tenth maximum [FWTM] = 6.73 mm) and (FWHM = 3.22 mm; FWTM = 7.39 mm), respectively. Whereas, I-123 provided with ME collimator a lower resolution (FWHM = 4.89 mm; FWTM = 9.89 mm). The sensitivity recorded by Tc-99 m, Sm-153, and I-153 were (31.21 Cps/MBq), (10.16 Cps/MBq), and (51.22 Cps/MBq), respectively. Conclusion: Tc-99 m and Sm-153 give the best and generally similar imaging properties with LEHR. For I-123, the ME collimator helps lowering the influence of high-energy gamma rays.

Keywords: Geometric, imaging, penetration, resolution, scattering, sensitivity, SIMIND


How to cite this article:
Bouzekraoui Y, Bentayeb F, Asmi H, Bonutti F. Comparison of image quality of different radionuclides technetium-99m, samarium-153, and iodine-123. Indian J Nucl Med 2019;34:201-4

How to cite this URL:
Bouzekraoui Y, Bentayeb F, Asmi H, Bonutti F. Comparison of image quality of different radionuclides technetium-99m, samarium-153, and iodine-123. Indian J Nucl Med [serial online] 2019 [cited 2019 Jul 20];34:201-4. Available from: http://www.ijnm.in/text.asp?2019/34/3/201/260754




   Introduction Top


In nuclear medicine, the radionuclides technetium-99 m (Tc-99 m) and iodine-123 (I-123) are most commonly used in single-photon emission computed tomography imaging. While, Samarium-153 (Sm-153) emits gamma radiation at 103 keV and has a physical half-life of 46.7 h, which is well suited for scintigraphic imaging.[1] Whereas, Sm-153 possesses both therapeutic beta and diagnostic gamma radiations, making it an alternative radionuclide to Y-90 in liver cancer treatment.[2] The Sm-153 imaging was useful in gastrointestinal scintigraphy.[3] In addition, Sm-153 has been used in various scintigraphic studies.[4],[5] Sm-153 has average, and maximum beta particle ranges in water of 0.5 mm and 3.0 mm, respectively. It is undeniable that the image quality strongly depends on the types of detected photons in the photopeak window. This photon retching the detector were classified as: the geometric component (passed without any interaction inside the collimator), the penetration component (passed through septa without attenuation), or the scatter component (scattered in the septa). In addition, photons that are absorbed in the collimator septa produce an X-ray component by the photoelectric effect.[6] Only the first component provides the correct information. The penetration and scattering effects in collimator degrade contrast, resolution, and quantification. As well as, image quality and quantification accuracy are affected by these factors.[7] Gamma camera cannot classify the detected photons into geometric, penetrated, or scattered photons. To solve this difficulty, we used Monte Carlo simulation technique to assess the geometric, penetration, and scatter contribution inside the photopeak window.[8] Whereas there are few studies related with the assessment of geometric, penetration, and scatter components.[9],[10],[11],[12]

The aim of this study was to evaluate the image quality for each radionuclide. We model the Siemens Symbia medical system Gamma Camera, using Monte Carlo SIMIND code.[13],[14],[15],[16],[17] Moreover, evaluate the geometric, scattered, and septal penetration components. We also evaluate the resolution and sensitivity (Cps/MBq).


   Materials and Methods Top


Images were acquired on a Siemens Medical System Symbia using both a medium-energy (ME), low-energy high resolution (LEHR), and ME collimators [Table 1]. I-123 was imaged with ME collimator, whereas Tc-99 m and Sm-153 were imaged with LEHR [Table 2]. Therefore, a point source has been placed at 12 cm from the detector surface and located in horizontal cylindrical water phantom of dimension 16 cm × 22 cm × 22 cm. We have modeled the detector with a radius equal to 25 cm having 2.54 cm NaI (Tl) crystal thickness. We made the intrinsic spatial resolution of 0.34 cm and energy resolution of 8.80% at 140 keV. We have used, the SIMIND Monte Carlo code developed by Michael Ljungberg.
Table 1: Collimators specifications

Click here to view
Table 2: Energies and intensities of gamma rays emitted from: The technetium-99m, samarium-153, and iodine-123 sources

Click here to view


Images were acquired using a 128 × 128 matrix and a pixel size = 0.34 mm × 0.34 mm. The images created by SIMIND in ImageJ software, National Institutes of Health, and the Laboratory for Optical and Computational Instrumentation (University of Wisconsin).[18]

[Table 2] shows photons energies and intensities of I-123, Tc-99 m, and Sm-153 radionuclides decay. For I-123 and Sm-153, ten peaks have been simulated, and three peaks have also been simulated for Tc-99 m [Table 2]. Simulations were carried out for two separate conditions: in air and with scattering medium (cylinder of water). In this simulation study, the main energy window was centered on the main energy peak of radionuclides for width of 20% of this energy (I-123 [159 keV]: 143–175 keV, Tc-99 m [141 keV]: 131–151 keV, and Sm-153 [103 keV]: 93–113 keV). The point spread function (PSF) was studied for each radionuclide. The full width at half maximum (FWHM) and the full width at tenth maximum (FWTM) were computed from the PSF to quantify the resolution. All imaging parameters have been compared with or without a scattering medium.


   Results and Discussion Top


The detected spectrum of radionuclides without collimator is shown in [Figure 1]. The energies are shown the energies of the main emission peaks of the isotopes.
Figure 1: Energy spectra of photons on the detector without the use of collimator

Click here to view


The different fractions detected within 20% peak energy window for all radionuclides are shown in [Table 3]. In water, the geometric fraction is large for Tc-99 m compared to the two other radionuclides. Whereas, in air this fraction, it is approximately the same for the three radionuclides. Because, these components depend on the energy of photons, object under study and collimator design parameters.
Table 3: Geometric, penetration, and scatter fractions achieved with the isotopes

Click here to view


[Figure 2] shows images of a point source for different radionuclides in water and air. The fogginess in the images formed by I-123 in water and air caused by scatter and penetration components due to high-energy gamma ray's emissions. The images formed by Sm-153 and TC-99 m show clearly a good quality.
Figure 2: Planar images created at the end of each simulation

Click here to view


Resolution and sensitivity (Cps/MBq) study

The PSF obtained for all of the isotopes, in air and water, are shown in [Figure 3]. Tc-99 m and Sm-153 offer a better resolution than the I-123 in both cases. The curve of I-123 with the ME collimator shows the effects of the septa which lowers the resolution.
Figure 3: Point spread function for technetium-99 m, Samarium-153, and Iodine-123

Click here to view


Images quality depends on resolution and also by sensitivity of the detector collimator system. [Figure 4] shows the comparison of sensitivity for three radionuclides.
Figure 4: Sensitivity at the end of simulation

Click here to view


[Figure 4] shows that for each isotope the water phantom decreases the sensitivity. For I-123, the ME produced the high sensitivity compared with LEHR for two other radionuclides.

Quantification of the resolution

FWHM and FWTM were calculated from profiles through the source images.

FWHM and FWTM are shown in [Table 4]; it shows that, for each isotope, the water phantom increases FWTM, while FWHM remains roughly the same. FWHM was better for the LEHR then for the ME collimator. In addition, FWHM was similar for both Tc-99 m and Sm-153.
Table 4: Full width at half maximum, full width at tenth maximum, and sensitivity data for the different radionuclides point source

Click here to view



   Conclusion Top


The effect of penetrated and scattered photons has been studied both qualitatively and quantitatively for Sm-153 and Tc-99 m and I-123. Since the results of the simulation show that the Sm-153 and Tc-99 m produced, generally, the best and similar imaging characteristics with LEHR collimator. However, the sensitivity obtained, with air and water phantom, was more than that of the other one. For I-123 imaging, the use of an ME collimator is important when the resolution is not required.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Tse JW, Wiebe LI, Noujaim AA. High specific activity [samarium-153] EDTA for imaging of experimental tumor models. J Nucl Med 1989;30:202-8.  Back to cited text no. 1
    
2.
Hashikin NA, Yeong CH, Abdullah BJ, Ng KH, Chung LY, Dahalan R, et al. Neutron activated samarium-153 microparticles for transarterial radioembolization of liver tumour with post-procedure imaging capabilities. PLoS One 2015;10:e0138106.  Back to cited text no. 2
    
3.
Yeong CH, Abdullah BJ, Ng KH, Chung LY, Goh KL, Sarji SA, et al. Production and first use of 153SmCl3-ion exchange resin capsule formulation for assessing gastrointestinal motility. Appl Radiat Isot 2012;70:450-5.  Back to cited text no. 3
    
4.
Marvola J, Kanerva H, Slot L, Lipponen M, Kekki T, Hietanen H, et al. Neutron activation-based gamma scintigraphy in pharmacoscintigraphic evaluation of an egalet constant-release drug delivery system. Int J Pharm 2004;281:3-10.  Back to cited text no. 4
    
5.
Wilding IR, Hardy JG, Sparrow RA, Davis SS, Daly PB, English JR, et al. In vivo evaluation of enteric-coated naproxen tablets using gamma scintigraphy. Pharm Res 1992;9:1436-41.  Back to cited text no. 5
    
6.
De Vries DJ, Moore S. Approximation of approximation of hexagonal holes by square holes in Monte Carlo simulation of gamma-camera collimation. IEEE Trans Nucl Sci 2002;49:2186-95.  Back to cited text no. 6
    
7.
Shafaei M, Ay MR, Sardari D, Dehestani N, Zaidi H. Monte Carlo assessment of geometric, scatter and septal penetration components in DST-XLi HEGP collimator. IFMBE Proc 2008;22:2479-82.  Back to cited text no. 7
    
8.
Ljungberg M, Larsson A, Johansson L. A new collimator simulation in SIMIND based on the delta scattering technique. IEEE Trans Nucl Sci 2005;52:1370-5.  Back to cited text no. 8
    
9.
Cot Sanz A, Pareto D, Sempau J, Bullich S, Pavia J, Calviño Tavares F, et al, Evaluation of the geometric, scatter, and septal penetration components in fan-beam collimators using Monte Carlo simulation. IEEE Trans Nucl Sci 2002;49:12-6.  Back to cited text no. 9
    
10.
Vandenberghe S, Van Holen R, De Beenhouwer J, Staelens S, Lemahieu I. Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131). Cancer Biother Radiopharm 2007;22:423-30.  Back to cited text no. 10
    
11.
Autret D, Bitar A, Ferrer L, Lisbona A, Bardiès M. Monte Carlo modeling of gamma cameras for I-131 imaging in targeted radiotherapy. Cancer Biother Radiopharm 2005;20:77-84.  Back to cited text no. 11
    
12.
Pandey AK, Sharma SK, Karunanithi S, Kumar P, Bal C, Kumar R. Characterization of parallel-hole collimator using Monte Carlo simulation. Indian J Nucl Med 2015;30:128-34.  Back to cited text no. 12
[PUBMED]  [Full text]  
13.
Ljungberg M. The SIMIND Monte Carlo Program Home Page. Available from: https://www.msf.lu.se/forskning/the-simind-monte-carlo-program. [Last accessed on 2019 Feb 10].  Back to cited text no. 13
    
14.
Rong X, Du Y, Ljungberg M, Rault E, Vandenberghe S, Frey EC. Development and evaluation of an improved quantitative (90) Y bremsstrahlung SPECT method. Med Phys 2012;39:2346-58.  Back to cited text no. 14
    
15.
Rong X, Frey EC. A collimator optimization method for quantitative imaging: Application to Y-90 bremsstrahlung SPECT. Med Phys 2013;40:082504.  Back to cited text no. 15
    
16.
Bahreyni Toossi MT, Islamian JP, Momennezhad M, Ljungberg M, Naseri SH. SIMIND monte carlo simulation of a single photon emission CT. J Med Phys 2010;35:42-7.  Back to cited text no. 16
    
17.
Roshan HR, Mahmoudian B, Gharepapagh E, Azarm A, Pirayesh Islamian J. Collimator and energy window optimization for 90Y bremsstrahlung SPECT imaging: A SIMIND monte carlo study. Appl Radiat Isot 2016;108:124-8.  Back to cited text no. 17
    
18.
Ferreira T, Rasband W. ImageJ Program. Available from: https://www.imagej.nih.gov/ij/download.html. [Last accessed on 2019 Feb 10].  Back to cited text no. 18
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
    Results and Disc...
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed114    
    Printed3    
    Emailed0    
    PDF Downloaded32    
    Comments [Add]    

Recommend this journal