ORIGINAL ARTICLE
Year : 2012  |  Volume : 27  |  Issue : 4  |  Page : 243-245  

Radiation surveillance in and around cyclotron facility


Department of Nuclear Medicine and PET, Post-Graduate Institute of Medical Education and Research, Chandigarh, India

Date of Web Publication23-Jul-2013

Correspondence Address:
Amandeep Kaur
#202, 'U' Block, Medical Campus, Guru Gobind Singh Medical College and Hospital, Faridkot - 151 203, Punjab
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-3919.115395

Rights and Permissions
   Abstract 

The cyclotron is the most widely used particle accelerator for producing medically important radio nuclides. Many medical centers in India have installed compact medical cyclotrons for on-site production of short-lived positron-emitting radio nuclides such as 18 F, 13 N, and 11 C. A mandatory requirement for cyclotron installation is radiation control permit from Atomic Energy Regulatory Board. Cyclotron radiation survey is an integral part of the overall radiation safety in the cyclotron facility. Radiation surveillance in and around a newly installed cyclotron was performed using ionization chamber counter and Geiger Muller counter before, during and after operating the cyclotron. The readings were recorded at various locations where a high radiation field was expected. The results were recorded, tabulated and analyzed. The highest exposure level (0.93 μSv) was found at the back wall of the radiochemistry lab facing the cyclotron vault. Reason for the high exposure of 0.93 μSv/h: Synthesis of 18 F-Fluoro-Deoxy-Glucose ( 18 F-FDG) was going in the synthesis module and activity ( 18 F) was present in the synthesis module when reading was taken. All other values were found to be below the recommended levels of exposure.

Keywords: Area survey, cyclotron, radiation surveillance


How to cite this article:
Kaur A, Sharma S, Mittal B R. Radiation surveillance in and around cyclotron facility. Indian J Nucl Med 2012;27:243-5

How to cite this URL:
Kaur A, Sharma S, Mittal B R. Radiation surveillance in and around cyclotron facility. Indian J Nucl Med [serial online] 2012 [cited 2019 Dec 16];27:243-5. Available from: http://www.ijnm.in/text.asp?2012/27/4/243/115395


   Introduction Top


The cyclotron is the most widely used particle accelerator for the production of Positron Emission Tomography (PET) radioisotopes (such as 18 F, 13 N, and 11 C). [1],[2],[3] An accelerator is a device that produces a well-defined, high-energy beam of charged particles with the high-beam intensity. [4] This beam strikes the target and leads to the production of radionuclide. During routine cyclotron operation and target bombardment beam loss may occur when the de-focused or misdirected particle beam inadvertently strikes the internal wall of the beam tube, can produce high levels of prompt neutrons and gamma rays. [5] The prompt fast neutrons slow down via multiple scattering events within the shielding and 10% are reflected back toward the cyclotron, thereby activating the various cyclotron components (including the beam tube, focus in g magnets, and ducts and piping). In addition, some prompt gamma and neutron radiations escape to the external environment through the shielding, producing direct external exposure. The neutrons, which penetrate the vault roof travel rather long distances and undergo multiple scattering events with nitrogen and oxygen in air, contributing to exposure via this "sky shine" radiation. With an adequate shielding; however, sky shine radiation can be reduced to negligibly low levels. [6],[7]

The major consideration in a cyclotron installation is the choice between a self-shielded and a vault-shielded device. [7] A self-shielded cyclotron may also require a vault, though one with thinner walls than those of a non-self-shielded cyclotron. [3],[8] In the former, retractable shielding is integrated into the cyclotron unit; in the latter, shielding is separate, and is built into the structure housing the device.

Still during the routine operation of cyclotron, it is mandatory to measure the radiation levels in and around the cyclotron. This helps to keep the radiation exposure to a level of minimum and also helps the radiological safety officer (RSO) and medical physicist (MP) to identify the high radiation level around cyclotron. [9],[10],[11],[12]

Routine cyclotron survey is an important part of overall radiation safety in the cyclotron facility. Facilities operating cyclotrons have radiological hazards that necessitate the implementation of an integrated radiation safety program.

This procedure should be routinely performed by the MP during the operation of the cyclotron. This helps MP/RSO to ensure the radiation safety of visitor, working staff, and radiation workers. RSO/MP can also tell whether the exposure is within the atomic energy regulatory board (AERB) permissible limit or not. [13] It can prevent unjustifiable radiation exposure to occupational worker and patients. To measure the radiation levels, every Nuclear Medicine Department should have radiation monitors, such as geiger muller (GM) counter and ionization chamber based survey meters. [14],[15],[16] Radiation monitors are instruments, which are designed to measure dose rate (in mSv/h), integrated dose (mSv) or contamination level (Bq/cm 2 ). The choice of instrument is governed by the nature and the level of radiation anticipated in the environment. [17] The present study is designed to survey the radiation in and around cyclotron area during its operation and to locate the area having high radiation exposure and at what time during the running of the cyclotron. The readings were compared using GM based and ionization chamber based survey meter. The background values calculated according to public annual exposure (1 mSv/h) according to AERB gives us 1000 μSv/year ≈ 365 days. 3 μSv/day and a working time of 12 h give exposure level of 0.25 μSv/h. [13]


   Materials and Methods Top


A new installed self-shielded cyclotron was area surveyed using GM and ionization chamber based survey meters. Radiation surveillance was performed before, during and after the cyclotron operation, using ionization chamber (INOVISION, USA) and GM counter (RAMGENE-1, ROTEM, USA) based detectors. The various areas in and around cyclotron are shown in the [Figure 1].
Figure 1: Layout of cyclotron showing different areas where radiation survey was carried out

Click here to view


Readings were taken with Ionization based survey meter and GM counter based survey meter continuously for 30 days at eight different areas.

Cyclotron facility was GE PET TRACER4.

Duration of the bombardment was average of 39 min/day.

Average of 1393 mCi/day activity of 18 F produced was produced in cyclotron.

Synthesis module was GE TRACER lab MXFDG.

Before the procedure for survey was started, the following preparations were performed. Firstly batteries of survey meters were checked. The survey meter was checked with a known radiation source. The meter was set at zero in a low-background radiation area. The type, Department of Energy (DOE) property number and calibration due date on the radiation survey form was recorded before the use of the survey meter. The surveyor wears a dosimeter before starting the procedure.

Different steps were performed during the radiation survey procedure. Radiation survey was carried out in all the locations possible to cover according to the layout and was carried out using GM based survey meter and ionization chamber based survey meter, simultaneously. Survey meter was held at arm's length, minimize exposure to the body. Radiation levels were monitored around PET trace4 self-shielded cyclotron (before bombardment, during bombardment, after bombardment) hot lab (HL) wall, console, vault door, vault wall (VW), outer wall, chillier side wall, quality control, room wall, and MP room. The results were recorded and tabulated. Appropriate statistical analysis was performed on the recorded readings. [18] Care was taken that survey does not take more than a few minutes.


   Results and Discussion Top


The details of the exposure around cyclotron using both types of survey meters are shown in the [Table 1]. Highest exposure level found in the HL after the bombardment was (0.93 μSv/h and 0.65 μSv/h by ionization chamber and GM counter. This may be due to the on-going synthesis in the hot cell. However, the second highest value (0.22 μSv/h) was found at VW and chiller side using ionization counter during bombardment. As soon as the bombardment was over, the exposure rate reduced. The [Figure 2] and [Figure 3] shows the average values before, during and after the bombardment using both ionization chamber and GM counter in μSv/h. Remaining values were within prescribed limit.
Figure 2: The average values with GM survey meter in μSv/h

Click here to view
Figure 3: The average values with Ionization based survey meter in μSv/h

Click here to view
Table 1: The average exposure values with GM counter and ionization counter in μSv/h at various locations

Click here to view


The graph [Figure 4] suggests positive correlation between the readings of GM counter and ionization chamber based survey meters, but is not in a perfectly linear correlation. The Spearman rank correlation coefficient between the readings of two survey meters is (r) =0.324. Correlation 'r' value suggests that there is random variation between the readings of two survey meters.
Figure 4: Comparison correlation between the readings of GM based survey meter (μSv/h) and ionization chamber based (μSv/h) survey meter

Click here to view


The background values calculated according to Public Annual Exposure (1 mSv/h) according to AERB gives us 1000 μSv/year ≈ 365 days. 3 μSv/day and a working time of 12 h give exposure level of 0.25 μSv/h. We have compared reading with 0.25 μSv/h at all the locations and found that the exposure is below this level at all the time during the run of the cyclotron. The high exposure level in the cyclotron area is only during the bombardment period of 40 min. During this time period a worker staying near the VW or HL will not receive on an average of more than 1 μSv dose daily and approximately, 300 μSv extra dose yearly. These values are thus well within the AERB annual limits. [13]


   Conclusion Top


The radiation survey was successfully performed in and around the newly installed cyclotron. All the values are below recommended levels of exposure. This reflected that the exposure levels outside the shield are minimal, which means that there was no leakage of the radiation outside the vault and all the values are below the recommended levels of exposure. Positive linear correlation with random variation exists between the readings of two survey meters.

 
   References Top

1.Fowler JS, Wolf AP. Positrons emitter-labeled compounds: Priorities and problems. In: Phelps ME, Mazziotta JC, Schelbert HR, editors. Positron emission tomography and autoradiography, principles and application for the brain and heart. New York: Raven Press; 1986. p. 391-450.  Back to cited text no. 1
    
2.Michael EP. Electronic generators. Positron emission tomography: Molecular imaging and its biological applications. Springer-Verlag New York, Inc.; 2004. p. 217-70.  Back to cited text no. 2
    
3.Saha GB. Cyclotron and production of PET radionuclides. Basics of PET imaging physics, chemistry, and regulations. 2 nd ed. Springer-Verlag New York LLC; 2005. p. 99-110.  Back to cited text no. 3
    
4.Livingston MS. High-energy accelerators. 2 nd ed. New York: Interscience Publishers, Inc.; 1954. p. 265-80  Back to cited text no. 4
    
5.Technical Report Series No. 471. Cyclotron produced radionuclides: guidelines for setting up a facility. International atomic energy agency. Vienna 2009. p. 21-48.  Back to cited text no. 5
    
6.White D. PET/CT and PET cyclotrons: Technical features and regulatory considerations. Frankfort, KY: Conference of radiation control program directors, Inc.; CRCPD national symposium on fusion imaging and multimodalities; 2004.  Back to cited text no. 6
    
7.Mukherjee B, Sartori E. A radiological safety and health physics database for cyclotrons accelerating protons and deuterons. Paris: Nuclear energy agency (NEA)/Organisation for economic co-operation and development (OECD); 2004.  Back to cited text no. 7
    
8.Pant GS, Senthamizhchelvan S. Initial experience with an 11 MeV self-shielded medical cyclotron on operation and radiation safety. J Med Phys 2007;32:118-23.  Back to cited text no. 8
[PUBMED]  Medknow Journal  
9.Hendee WR, Ritenour ER. Protection from external sources of radiation. Medical imaging physics. 4 th ed. 2002 by Whilay-Liss, Inc, New York. p. 435-54.  Back to cited text no. 9
    
10.Zanzonico P, Dauer L, St Germain J. Operational radiation safety for PET-CT, SPECT-CT, and cyclotron facilities. Health Phys 2008;95:554-70.  Back to cited text no. 10
    
11.Russo AA, Ferrari P, Casale M, Delia R. The radioprotection management of a PET department with a cyclotron and radiopharmacy laboratory, in accordance with Italian legislation. Radiat Prot Dosimetry 2011;147:240-6.  Back to cited text no. 11
    
12.Mollah AS. Regulation and radiation safety issues of medical cyclotron facility: An overview, Bangladesh. J Nucl Med 2008;11-61.  Back to cited text no. 12
    
13.Shaprio Jacob. Principal of protection against ionization radiation. Radiation Protection: A guide for scientists regulators, and physicians. 4 th ed.p.cm. La Editorial. UPR 2002.p. 18-166.  Back to cited text no. 13
    
14.Kieran M. Gas-filled radiation detectors. Basic physics of nuclear medicine. Wikibooks 2006. p. 57-62.  Back to cited text no. 14
    
15.Padhy AK, Solanki KK. Instrumentation. Nuclear medicine resources manual. International atomic energy agency, Vienna; Austria 2006.p. 107-64.  Back to cited text no. 15
    
16.Saha GB. Instruments for radiation detection and measurement. Fundamentals of Nuclear Pharmacy. 5 th ed. Springer-Verlag; New York, Inc; 2004. p. 50-3.  Back to cited text no. 16
    
17.Podgorsak EB. Radiation monitoring instruments. Radiation oncology physics: A handbook for teachers and students. international atomic energy agency, Vienna; Austria 2005. p. 108-12.  Back to cited text no. 17
    
18.Currell G, Dowman A. Essential mathematics and statistics for science. 2 nd ed. United Kingdom: Willey-Blackwell Publication; 2009. p. 315-24.  Back to cited text no. 18
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Distribution of residual long-lived radioactivity in the inner concrete walls of a compact medical cyclotron vault room
Toshioh Fujibuchi,Akihiro Nohtomi,Shingo Baba,Masayuki Sasaki,Isao Komiya,Yoshiyuki Umedzu,Hiroshi Honda
Annals of Nuclear Medicine. 2014;
[Pubmed] | [DOI]
2 Radiation accidents and their management
Jamshed B. Bomanji,Fuad Novruzov,Sobhan Vinjamuri
Nuclear Medicine Communications. 2014; 35(10): 995
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
    Results and Disc...
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1535    
    Printed43    
    Emailed1    
    PDF Downloaded117    
    Comments [Add]    
    Cited by others 2    

Recommend this journal